Merlin can install and configure your entire Ozone Precursor System from its original cartons. The entire process takes only 3-5 days with all the necessary gases and power requirements available. We recommend each site have two phone lines available for voice and modem connections for the standard installation or 3 phone lines if hourly data will be transmitted. For systems requiring hourly transmission, Merlin provides a second communications terminal. This isolates transmission activities from the system computer and reduces the potential for data losses due to transmission problems on the communications terminal.

In addition, based on our experience with the power requirements of the system, Merlin recommends at least 2 separate 25 amp circuits be available for the Gas Chromatograph and ATD. Separate 10 amp circuits should be provided for all ancillary equipment. Air Conditioning and other heavy equipment should be well grounded and isolated from the system to reduce line noise, which can adversely affect the electrometer signals.

Custom installation and configuration provides each field site with a complete monitoring system, which can be configured and tested as a complete system. Software from several sources and hardware from multiple vendors can all be integrated into an effective system from the initial installation. Miscellaneous site-specific problems can be effectively managed from the onset providing a problem free monitoring site.

 

 