Merlin
can install and configure your entire Ozone Precursor System from
its original cartons. The entire process takes only 3-5 days with
all the necessary gases and power requirements available. We recommend
each site have two phone lines available for voice and modem connections
for the standard installation or 3 phone lines if hourly data
will be transmitted. For systems requiring hourly transmission,
Merlin provides a second communications terminal. This isolates
transmission activities from the system computer and reduces the
potential for data losses due to transmission problems on the
communications terminal.
In
addition, based on our experience with the power requirements
of the system, Merlin recommends at least 2 separate 25 amp circuits
be available for the Gas Chromatograph and ATD. Separate 10 amp
circuits should be provided for all ancillary equipment. Air Conditioning
and other heavy equipment should be well grounded and isolated
from the system to reduce line noise, which can adversely affect
the electrometer signals.
Custom
installation and configuration provides each field site with a
complete monitoring system, which can be configured and tested
as a complete system. Software from several sources and hardware
from multiple vendors can all be integrated into an effective
system from the initial installation. Miscellaneous site-specific
problems can be effectively managed from the onset providing a
problem free monitoring site.